User’s and Service Guide 5
87050A Option K22 Preparations
Preparations
1. Ensure that the "Meeting Electrical and Environmental Requirements" on page 3 are
met.
2. Verify that the power cable is not damaged, and that the power source outlet provides a
protective earth ground contact. Note that the Figure 2 depicts only one type of power
source outlet. Refer to Figure 3 on page 6 to see the different types of power cord plugs
that can be used with your test set. Cables are available in different lengths. For
descriptions and part numbers of cables other than those described in Figure 3, Refer to
"Contacting Agilent" on page 74.
3. If this product is to be powered by autotransformer, make sure the common terminal is
connected to the neutral (grounded) side of the ac power supply.
WARNING This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only
be inserted into a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor, inside or
outside the instrument, is likely to make the instrument dangerous.
Intentional interruption of the protective conductor is prohibited.
Figure 2 Protective Earth Ground
CAUTION Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.