Vue de la page 73
66 User’s and Service Guide
87050A Option K22 Theory of Operation
Theory of Operation
The theory of operation begins with a general description of the Agilent 87050A Option
K22 multiport test set. This is followed by more detailed operating theory. The operation of
each group is described briefly, to the assembly level only. Detailed component-level circuit
theory is not provided.
System Theory
The test set consists of three main components: a power supply, front panel display, and
main switch board. The purpose of the power supply is to supply power to both the front
panel display and the main switch board. The front panel display serves to indicate the
switching paths to the user. Finally, the main switch board does the actual switching
between the different ports.
A1 Power Supply Theory
The switching power supply provides regulated dc voltages to power all assemblies in the
test set. A dc cable provides power to the main switch board. A connector from the main
switch board to the front panel display provides dc power and control signals to the front
panel. The power supply provides the following supplies: +5V, +24V, +12V.
The power LED on the front panel indicates that the instrument is on and that the power
supply is providing power.
A2 Front Panel Display Theory
The front panel display consists of an LCD. The LCD is divided into two lines, forward and
reverse. The first line indicates which of the twelve ports are connected to the forward
path. The second line indicates which of the twelve ports are connected to the reverse path.
Control signals and DC power are provided via a cable connected to the main switch board.
A3 Controller (Mother Board) and Switch Driver (Daughter Board)
Board Theory
Refer to Figure 16 on page 65 for the following discussion.
The mother and daughter boards provide the bias for the switching paths for the various
ports to the A i/p or B i/p ports. The front panel display contains an LCD that indicates the
switched ports. A particular test port (1 through 12) can be in one of three states. The three
states are:
1. Switched to the forward path
2. Switched to the reverse path
3. Terminated in 50 Ω